Thiết bị mô phỏng tín hiệu bệnh nhân
Model: AMPS-1
Nhà sản xuất: Datrend / Canada
Các chức năng chính:
ECG General:
Full 12-Lead ECG with 9 independent outputs for each signal lead referenced to RL.
Output Impedances: 500, 1000, 1500, & 2000 ohms to RL.
High Level Output: 0.5 V/mV of low level selection.
Amplitude Accuracy: ± 2% 2 Hz Square Wave (Lead II).
Normal Sinus Rhythm:
Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 BPM.
Accuracy ± 1%.
Amplitudes (Lead II): 5mV, 4mV, 3mV, 2mV, 1mV, .5mV, .25mV, .1mV.
ST Segments: 16 total – 8 elevated & 8 depressed.
ST Segment Levels (Lead II): -0.8 mV to +0.8 mV in 0.1 mV steps on Lead II, .
Axis Deviation:
Normal (intermediate), horizontal, and vertical. Modifies baseline ECG during arrhythmias. Neonatal Mode: ECG R wave width is reduced to 40ms.
ECG Performance Testing:
Square Wave: 2 Hz
Square Wave: 0.125 Hz
Pulse: 4.0 secs
Sine Waves: 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz.
Triangle Wave: 2 Hz
R Wave Detector Test: 60 BPM haver-triangle wave with selectable ampl. and width.
Width: 8.0 ms to 200 ms (12 selections)
Amplitude (Lead II and V Leads): 5mV to 0.5mV.
Pacemaker:
Asynchronous
Demand with frequent sinus beat
Demand with occasional sinus beat
A-V sequential
Non-capture non-function
Rhythms: Async 75 BPM, Demand 1, Demand 2, AV Seq,
Non capture, Non function.
Pulse: -700 mV to +700 mV.
Accuracy : 5%
Width: 0.1, 0.2, 0.5, 1.0, 2.0 ms. Accuracy is 5%.
Pulse Polarity: Positive or negative.
Synchronization:
Accepts input during defib training.
RS-232 Interface:
RS-232 interface to PC.
Defibrillator Training:
With two emergency scenarios and a cardioversion procedure, AMPS-1 can be used for basic defibrillator training.
Cardiac Output:
Built in Cardiac Output feature, activation optional 4 adjustable injectate temperature selections that are factory set and user adjustable
Baseline of 36, 37 and 38 degrees °C.
Selections for 2 and 20 degrees °C.
Cardiac Output Selections:
Faulty Injectate Curve
Left to Right Shunt Curve
C.O. of 3, 4, 4.5, 5, 5.5, 6, 6.5, 7 l/min
Cal Pulse: 1o for 1 second
Cal Pulse: 1o for 4 seconds
CC .561 for 2 degrees injectate
CC .608 for 20 degrees injectate